Every moving part below is driven by a one-dimensional physical model of the haemofilter. Change the prescription and watch pressures, haematocrit, oncotic pressure and solute clearance respond.
View
The circuit
Particle speed in each line is proportional to its flow rate. The share of red dots in a blood line equals its haematocrit. Tap any part of the circuit, or a button below, to highlight it.
Highlight
Red cellsPlasmaReplacement fluidUltrafiltrate (effluent)
Prescription and patient
Effluent rate = replacement rate + net fluid removal: in CVVH the machine sets the effluent pump, and filter pressures settle wherever they must to deliver it.
Inside one hollow fibre
A longitudinal slice through a single 200 µm fibre. Blood moves with a parabolic (Poiseuille) velocity profile. Water and small solutes are dragged through the membrane where local filtration occurs; cells and proteins are retained.
Red cellPlasma proteinWaterSolute
Speed
Along the filter
Steady-state profiles from blood inlet (0 cm) to outlet (25 cm).
Clearance and dose
The mathematics, with current values
Assumptions and limits of the model
Steady state, one-dimensional: the filter is divided into 120 axial segments, and all fibres are assumed to behave identically (no flow maldistribution).
Fibre geometry is fixed at 200 µm internal diameter and 25 cm length; fibre number is derived from surface area. Real filters vary.
Plasma proteins are fully reflected (σ = 1) for oncotic purposes; the albumin sieving figure is used only to estimate losses.
Membrane permeability is a single lumped value. Protein adsorption, the secondary membrane, concentration polarisation and progressive clotting are represented only by lowering it; set it low to model a fouled filter.
Solutes are tracked in plasma water only. Movement of urea between red cells and plasma, protein binding (beyond what the sieving coefficient already reflects) and solute in replacement fluid are ignored.
Effluent-side pressure is uniform; no diffusion occurs (pure convection, no dialysate); temperature effects on viscosity are ignored.
Access pressure is a rough linear estimate; in reality it depends on catheter size, position and patient factors.
The 24-hour projection is single-pool with no generation, so it shows the shape of removal, not a predicted lab value.
Sieving coefficients are typical literature values for high-flux synthetic membranes and vary between membranes and studies.
Lactate: one compartment with constant production and a fixed endogenous clearance. Real lactate metabolism is saturable and falls in liver failure and shock. Replacement fluid is assumed to be bicarbonate-buffered and lactate-free.
Phosphate: the pseudo-one-compartment model was derived in maintenance haemodialysis, not critical illness, and assumes the mobilisable tissue pool never runs down. Over days of CVVH, body stores do fall, and without supplementation hypophosphataemia develops.
An educational model for building intuition. It is not a prescribing tool.
Sources
Pries AR, Neuhaus D, Gaehtgens P. Blood viscosity in tube flow: dependence on diameter and hematocrit. Am J Physiol. 1992;263:H1770–H1778.
Landis EM, Pappenheimer JR. Exchange of substances through the capillary walls. In: Handbook of Physiology, Circulation, vol. 2. American Physiological Society; 1963.
Levraut J, Ciebiera JP, Jambou P, et al. Effect of continuous venovenous hemofiltration with dialysis on lactate clearance in critically ill patients. Crit Care Med. 1997;25:58–62.
Troyanov S, Cardinal J, Geadah D, et al. Solute clearances during continuous venovenous haemofiltration at various ultrafiltration flow rates using Multiflow-100 and HF1000 filters. Nephrol Dial Transplant. 2003;18:961–966.
Agar BU, Akonur A, Lo YC, Cheung AK, Leypoldt JK. Kinetic model of phosphorus mobilization during and after short and conventional hemodialysis. Clin J Am Soc Nephrol. 2011;6:2854–2860.
Leypoldt JK, Agar BU, Akonur A, et al. Determinants of phosphorus mobilization during hemodialysis. Kidney Int. 2013 (HEMO Study analysis).
Sharma S, Waikar SS. Phosphate balance in continuous venovenous hemofiltration. Am J Kidney Dis. 2013;61:1043–1045.